miniPV®-X

Sku#

9725-C11

miniPV®-X Automated Dilute Solution/Kinematic Viscometer

In stock

This is non stock item.

miniPV-X is a fully automated, single-bath, benchtop viscometer with a 10-position sample handler for unattended processing and testing in organic solvents and aqueous solutions for dilute solution viscosity of polymers and kinematic viscosity for general purpose applications.

Viscosity range for dilute solution viscosity is 0.3 mm2/s (cSt) to 700 mm2/s (cSt)* and for kinematic viscosity 0.02 mm2/s (cSt) to 1,200 mm2/s (cSt)* at 15 °C to 100 °C.

Protect your investment and ensure your instrument continues to provide quality and reliable results for years to come with a CANNON Service Plan.

Factory installation of this unit is required. If factory installation is waived, only parts are covered during the warranty period.  Damage occurring from incorrect installation is not covered by warranty.

*Upper viscosity measurements may be limited by test temperature and sample type.

 

Designed to meet your viscosity testing needs

  • On-board software with specialized polymer calculations determines inherent, intrinsic, kinematic, reduced, relative, and specific viscosity, as well as, molecular weight
  • Integrated TE cooling provides superb temperature control from 15 °C to 100 °C
  • Dilute solution polymer viscosity range: 0.3mm²/s (cSt) to 700mm²/s (cSt); both single range and special dual range viscometer tubes are available*
  • Kinematic viscosity range: 0.02mm²/s (cSt) to 1,200mm²/s (cSt)*

Compact,robust design

  • Fits in roughly the same benchtop area as an analytical balance to conserve valuable lab space
  • Sealed system allows for measurement of atmospherically sensitive and volatile samples
  • Compatible with most aggressive aqueous, aqueous, caustic, halogenated, organic, and salt solutions

Fully automated benchtop testing

  • Software controls the instrument and facilitates tasks such as calibration, data entry, method specification, calculation selection, report formatting, and data exporting
  • Single PC manages up to 4 instruments using VISCPRO® software
  • Reduces operator to operator variability
  • Replaces manual tube cleaning with automated washing and drying

Simplified maintenance & test versatility

  • Modular bath for easy maintenance access
  • Operators physically replace tubes in minutes, eliminating the need to schedule related service calls
  • Single-point temperature calibration avoids need for tube recalibration and maximizes test flexibility
  • Standard dual solvent input

*Upper viscosity measurements may be limited by test temperature and sample type.

Specifications
Unit Weight

Unit: 24 kg (53 lb)
Power Supply: 11 kg (24 lb)
Waste Receiver: 6 kg (13 lb)

Method ASTM D445/D446, ASTM D1243, ASTM D1795, ASTM D2857, ASTM D3591, ASTM D4243, ASTM D5226, DIN 51562-1, IEC 60450, ISO 1628-1, ISO 1628-2, ISO 1628-4, ISO 1628-6, ISO 3104/3105, ISO 5351, Ph. Eur. 1472, Tappi T230/PAPTAC G.24P, USP 911
Dimensions (W x D x H)

Unit: 25.4 cm × 39.6 cm × 61.9 cm (10.0 in × 15.6 in × 24.375 in)
Power Supply: 33.0 cm × 39.6 cm × 17.2 cm (13.0 in × 15.6 in × 6.8 in)
Waste Receiver: 33.0 cm × 39.6 cm × 17.2 cm (13.0 in × 15.6 in × 6.8 in)

Shipping Dimensions (W x D x H) 73.7 cm × 63.5 cm × 94.0 cm (29.0 in × 25.0 in × 37.0 in)
Shipping Weight (With All Items) 117 kg (258 lb)
Maximum Throughput Cycle time as low as 12 minutes (includes washing and drying time)
Sample Capacity 10
Viscosity Range

Dilute Solution: 0.3 mm2/s (cSt) to 700 mm2/s (cSt)*
Kinematic: 0.02 mm2/s (cSt) to 1,200 mm2/s (cSt)*

*Upper viscosity measurements may be limited by test temperature and sample type.

Timing Resolution 0.01 s (timing accuracy to ± 0.001 s)
Temperature Range and Accuracy 15 °C to 100 °C, ± 0.01 °C
Minimum Sample/Solvent Volume 8 mL (as low as 1.6 mL with low-volume option) sample/24 mL solvent for wash
Operating Conditions 15 °C to 30 °C, 10% to 75% relative humidity (non-condensing), Installation Category II; Pollution Degree 2
Electrical Specifications 100 Vac, 50/60 Hz; 115 Vac, 50/60 Hz; 230 Vac, 50/60 Hz; 1000 watts power consumption
Compliance CE Mark; EMC directive (2004/108/EC); Low voltage directive (2006/95/EC); HI-POT (1900 Vdc, 60 sec.); ROHS
Data Output RS-232 via RS-485