miniPV® Automated Dilute Solution/Kinematic Viscometer
In stock
9725-A91
miniPV® Automated Dilute Solution/Kinematic Viscometer
In stock
miniPV is a fully automated, single-sample, benchtop viscometer for unattended processing and testing in organic solvents and aqueous solutions for dilute solution viscosity of polymers and kinematic viscosity for general purpose applications.
Viscosity range for dilute solution viscosity is 0.3 mm2/s (cSt) to 700 mm2/s (cSt)* and for kinematic viscosity 0.02 mm2/s (cSt) to 1,200 mm2/s (cSt)* at 15 °C to 100 °C.
Protect your investment and ensure your instrument continues to provide quality and reliable results for years to come with a CANNON Service Plan.
Factory installation of this unit is required. If factory installation is waived, only parts are covered during the warranty period. Damage occurring from incorrect installation is not covered by warranty.
*Upper viscosity measurements may be limited by test temperature and sample type.
*Upper viscosity measurements may be limited by test temperature and sample type.
Unit Weight | Unit: 24 kg (53 lb) |
---|---|
Method | ASTM D445/D446, ASTM D1243, ASTM D1795, ASTM D2857, ASTM D3591, ASTM D4243, ASTM D5226, DIN 51562-1, IEC 60450, ISO 1628-1, ISO 1628-2, ISO 1628-4, ISO 1628-6, ISO 3104/3105, ISO 5351, Ph. Eur. 1472, Tappi T230/PAPTAC G.24P, USP 911 |
Dimensions (W x D x H) | Unit: 25.4 cm × 39.6 cm × 61.9 cm (10.0 in × 15.6 in × 24.375 in) |
Shipping Dimensions (W x D x H) | 73.7 cm × 63.5 cm × 94.0 cm (29.0 in × 25.0 in × 37.0 in) |
Shipping Weight (With All Items) | 117 kg (258 lb) |
Maximum Throughput | Cycle time as low as 12 minutes (includes washing and drying time) |
Sample Capacity | 1 |
Viscosity Range | Dilute Solution: 0.3 mm2/s (cSt) to 700 mm2/s (cSt)* *Upper viscosity measurements may be limited by test temperature and sample type. |
Timing Resolution | 0.01 s (timing accuracy to ± 0.001 s) |
Temperature Range and Accuracy | 15 °C to 100 °C, ± 0.01 °C |
Minimum Sample/Solvent Volume | 8 mL (as low as 1.6 mL with low-volume option) sample/24 mL solvent for wash |
Operating Conditions | 15 °C to 30 °C,10% to 75% relative humidity (non-condensing),Installation Category II; Pollution Degree 2 |
Electrical Specifications | 100 Vac,50/60 Hz; 115 Vac,50/60 Hz; 230 Vac,50/60 Hz; 1000 watts power consumption |
Compliance | CE Mark; EMC directive (2004/108/EC); Low voltage directive (2006/95/EC); HI-POT (1900 Vdc, 60 sec.); ROHS |
Data Output | RS-232 via RS-485 |