CT-500 (240 V)

Sku#

9726-A15

CT-500 Constant Temperature Bath (240 V) 



In stock

The CT-500 series constant temperature bath (240 V) accommodates viscometer tubes requiring a bath depth up to 12" and provides reliable temperature control from 20 °C to 100 °C. The unit accommodates up to 7 simultaneous viscosity determinations.

For Kinematic Viscosity from 20 °C to 100 °C. Conforms to test methods ASTM D445/D446 and ISO 3104/3105. 

Precise and accurate temperature control

  • Temperature range: 20 °C to 100 °C ± 0.01 °C*
  • Temperature selection is variable between 20 °C and 100 °C or selectable at ten pre-set temperatures (20 °C, 25 °C, 30° C,
    40 °C, 50 °C, 60 °C, 70 °C, 80 °C, 90 °C, and 100 °C)
  • Meets ASTM D445 temperature sensitivity requirements for kinematic viscosity determination
  • Motor driven stirrer ensures a uniform bath temperature

Safe operation

  • Bath thermistor senses over-temperature fault and removes power
  • Thermistor disconnection results in removal of bath heater power
  • Liquid level sensor prevents operation if bath liquid is not at a safe operating level

Efficient, user-friendly design

  • Seven glass viscometer ports (each 2” in diameter) and two thermometer ports (each 0.4” in diameter) allow up to 7 simultaneous viscosity determinations**
  • Two electric heating elements quickly heat bath media to the selected temperature
  • Stainless steel baffle provides a plain reflective viewing background
  • Convenient drawer provides easy access to electrical controls

Reliable performance

  • Well tested CANNON bath platform offers reliability and dependable support
  • All wetted components are manufactured from high quality stainless steel or glass
  • The heavy duty aluminum frame offers a corrosion-resistant epoxy finish

* Test temperatures < 40 °C and high ambient temperature test environments require an optional bath cooler
** Other port configurations are available as a special order

Specifications
Unit Weight 22 kg (48 lb)
Dimensions (W x D x H) 40.7 cm × 36.2 cm × 61.0 cm (16 in × 14.25 in × 24 in)
Shipping Weight (With All Items) 36.3 kg (80 lb)
Sample Capacity 7
Temperature Range and Accuracy 20 °C to 100 °C, ± 0.01 °C*, * Test temperatures < 40 °C and high ambient temperature test environments require an optional bath cooler
Operating Conditions 15 °C to 30 °C, 10% to 75% relative humidity (non-condensing), Installation Category II; Pollution Degree 2
Electrical Specifications 120 Vac, 50/60 Hz; 240 Vac, 50/60 Hz; 1000 watts power consumption
Compliance CE Mark; EMC directive (2004/108/EC); Low voltage directive (2006/95/EC); HI-POT (1900 Vdc, 60 sec.); ROHS
Bath Volume 17 L (4.5 gal)
Bath Height 12 in