DVR-3000 W/ VACUUM PUMP 100-240V 50/60Hz

Sku#

9726-T70

DVR-3000 with Vacuum Pump 100-240V, 50/60Hz

In stock

The DVR-3000 digital vacuum regulator provides precise measurement and control of vacuum at 300 mm Hg below atmospheric pressure. Vacuum set point can be altered to any value within the operational range of 95 mm Hg to 450 mm Hg below atmospheric pressure. Solid state components contain no hazardous mercury.

Product Information

Precise Control of Vacuum

  • Operating range: 95—450 mm Hg below atmospheric pressure ± 0.5 mm Hg
  • Supplied with NIST certification of measurement accuracy
  • Vacuum level displayed in mm Hg
  • Preset to regulate vacuum at 300 mm Hg below atmospheric pressure (consistent with ASTM D2171 requirements)

Configurable to User-Specific Needs

  • Vacuum set point can be effortlessly changed to fit custom applications
  • Trusted CANNON® platform offers reliability and dependable support

Safe, Easy-to-Use Design

  • Solid state components do not contain hazardous mercury
  • Easy control input
  • Small footprint; 16.8 cm x 25.0 cm x 18.5 cm (6.62 in x 10 in x 7.25 in)
Specifications
Unit Weight 5 kg (11 lb)
Method ASTM D2171
Dimensions (W x D x H) 16.8 cm 25.0 cm x 18.5 cm (6.62 in x 10 in x 7.25 in)
Shipping Dimensions (W x D x H) 40.64 cm x 33.0 cm x 33.0 cm (16 in x 13 in x 13 in)
Shipping Weight (With All Items) 6.8 kg (15 lb)
Operating Conditions 15 °C to 30 °C, 10% to 75% relative humidity (non-condensing), Installation Category II; Pollution Degree 2
Electrical Specifications

100—240VAC, 50/60 Hz; 175 watts power consumption

Compliance

CE Mark; EMC directive (2004/108/EC); Low voltage directive (2006/95/EC); HI-POT (1900 VDC, 60 sec.); ROHS

Additional Standard Features
Vacuum Range & Accuracy

95—450 mm Hg below atmospheric pressure
± 0.5 mm Hg

Safety 746 mm Hg below atmospheric pressure