CMVC-RO-14 Cannon-Manning Vaccum Viscometer w/Round Holder Calibrated
9724-L80
Cannon-Manning Vaccum Viscometer
Size 14, Calibrated
With permanently attached round metal holder (for hole 51 mm (2 in) in diameter.
In stock
9724-L80
Cannon-Manning Vaccum Viscometer
Size 14, Calibrated
With permanently attached round metal holder (for hole 51 mm (2 in) in diameter.
In stock
The Cannon-Manning Vacuum Viscometer measures the viscosity of highly viscous materials according to ASTM D2171. Vacuum of 300 mm Hg is applied to small arm with timing bulbs; two bulbs are incorporated to extend viscosity range of viscometer.
Minimum sample volume is 6 mL. Minimum bath depth is 180 mm (7.1 in).
Viscometer is calibrated and is provided with a certificate of calibration and instruction sheet. See expanded uncertainty for more information.
Viscometer constant is the same at all temperatures. Approximate constant at 300 mm Hg vacuum is 200 poise/second for bulb B and 60 poise/second for bulb C.
Viscosity range is 3600–80000 poise.
Viscometer comes with a permanently attached round metal holder, for hole 51 mm (2 in) in diameter. Viscometer without attached holder is also available, see 9724-H80
Method | ASTM D2171 |
---|