CMVC-10 Cannon-Manning Vaccum Viscometer Calibrated

Sku#

9724-H68

Cannon-Manning Vaccum Viscometer, size 10, Calibrated

In stock

The Cannon-Manning Vacuum Viscometer measures the viscosity of highly viscous materials according to ASTM D2171. Vacuum of 300 mm Hg is applied to small arm with timing bulbs; two bulbs are incorporated to extend viscosity range of viscometer.

Minimum sample volume is 6 mL. Minimum bath depth is 180 mm (7.1 in).

Viscometer is calibrated and is provided with a certificate of calibration and instruction sheet. See expanded uncertainty for more information.

Viscometer constant is the same at all temperatures. Approximate constant at 300 mm Hg vacuum is 2 poise/second for bulb B and 0.6 poise/second for bulb C.

Viscosity range is 36–800 poise.

Holder sold separately. For holder, see 9726-N15. Viscometer with permanently attached round metal holder also available, see 9724-L68.

Specifications
Method ASTM D2171