CAV® 4.2
9725-B35
In stock
9725-B35
In stock
CAV 4.2 is a fully automated, dual-bath, benchtop viscometer with two 14-position sample handlers for unattended D445 processing. Each of the two tubes (one per bath) covers a 100-fold viscosity range at values between 0.5 mm2/s (cSt) and 10,000 mm2/s (cSt) from 15 °C to 150 °C (with available options).
Common Applications include formulated oil analysis, hydraulic oil analysis, additive analysis, marine fuel testing, base stock analysis, light and heavy fuel testing, waxes/paraffin, and crude oil testing.
Viscometer Tube Configuration Information
Protect your investment and ensure your instrument continues to provide quality and reliable results for years to come with a CANNON Service Plan.
Factory installation of this unit is required. If factory installation is waived, only parts are covered during the warranty period. Damage occurring from incorrect installation is not covered by warranty.
*Some upper viscosity measurements may be limited by test temperature and sample type. Ubbelohde-style tubes (for volatile products that are not a pure solvent and high melt waxes) are also available in 100-fold increments and fast run 10-fold increments from 0.5 mm²/s (cSt) to 10,000 mm²/s (cSt).
Unit Weight | 63.0 kg (140.0 lb) |
---|---|
Method | ASTM D445/D446, IP 71, ISO 3104/3105, SAE J300 |
Dimensions (W x D x H) | 36.0 cm × 66.0 cm × 72.0 cm (14.3 in × 26.0 in × 28.5 in) |
Shipping Dimensions (W x D x H) | 81.3 cm × 53.3 cm × 99.0 cm (32.0 in × 21.0 in × 39.0 in) |
Shipping Weight (With All Items) | 96.6 kg (213.0 lb) |
Maximum Throughput | 24 tests per hour |
Sample Capacity | 28 positions (2 x 14 positions), automated |
Viscosity Range | 0.5 mm²/s (cSt) to 10,000 mm²/s (cSt) in 100-fold increments (depending on viscometer tube selection). Fast run tubes are also available. |
Timing Resolution | 0.01 s (timing accuracy to ± 0.001 s) |
Temperature Range and Accuracy | 15 °C to 100 °C,± 0.01 °C (temperatures within 5 °C of ambient and below require bath cooling) |
Minimum Sample/Solvent Volume | 8 mL sample (as little as 3 mL with fast run tubes)/15 mL solvent per test |
Operating Conditions | 15 °C to 30 °C, 10% to 75% relative humidity (non-condensing),Installation Category II; Pollution Degree 2 |
Electrical Specifications | 100 Vac, 50/60 Hz; 115 Vac, 50/60 Hz; 230 Vac, 50/60 Hz; 1200 watts power consumption |
Compliance | CE Mark; EMC directive (2004/108/EC); Low voltage directive (2006/95/EC); HI-POT (1900 Vdc,60 sec.); ROHS |
Additional Standard Features | Internal heated waste drain lines,dual solvent input |
Data Output | USB |