TESC-5133

Sku#

9725-F82

TESC-5133 Low Temperature Viscometer with Brookfield® DV2T



In stock

The compact, economical TESC-5133 Low Temperature Viscometer integrates and automates the ASTM D5133 process of conditioning and testing low temperature viscosities of engine lubricants. The TESC 5133 Windows® 10 software application records sample temperature and viscosity throughout the test, then computes the temperature (Gel Index Temperature) at which there is a rapid change in viscosity (Gelation Index).

Common applications include gear oils, automatic transmission fluids, torque and tractor fluids, industrial and automotive hydraulic oils, industrial lubricants, and other fluids where low-temperature viscosity is critical.

Protect your investment and ensure your instrument continues to provide quality and reliable results for years to come with a CANNON Service Plan.

 

Fully integrated and automated conditioning and testing (−42 °C to +90 °C)

  • Provides an automated alternative to ASTM D5133 testing
  • Reduces result variability due to temperature fluctuation and operator intervention
  • Provides superior precision
  • Allows for unattended operation

Sample thermal conditioning completed without operator intervention

  • Reduces instrument set-up time
  • TESC 5133 Windows® 10 software application records sample temperature and viscosity throughout the test, then computes the temperature (Gel Index Temperature) at which there is a rapid change in viscosity (Gelation Index)

Cost-effective, ultra-compact design

  • Allows installation of multiple TESC Systems in a relatively small benchtop area for redundancy and greater test versatility
  • Conserves precious bench space

Easy-lift system

  • Configured for use with the Brookfield® DV2T viscometer 1/2 RV
  • Simplifies positioning of the viscometer head

Solid-state thermoelectric cooling

  • Peltier cooling is environmentally friendly and does not require the use of hazardous bath fluids
  • Enclosed sample chamber provides a uniform, controlled temperature environment during the entire test
Specifications
Unit Weight 19.5 kg (43 lb) including DV2T viscometer
Method ASTM D5133
Dimensions (W x D x H) 16.5 cm × 51.0 cm × 76.0 cm (6.5 in × 20.0 in × 30.0 in)
Shipping Dimensions (W x D x H) 73.7 cm × 71.1 cm × 58.4 cm (29.0 in × 28.0 in × 23.0 in)
Shipping Weight (With All Items) 34 kg (75 lb)
Sample Capacity 1
Temperature Range and Accuracy −42 °C to 90 °C (± 0.1 °C)
Minimum Sample/Solvent Volume 20 mL
Operating Conditions 15 °C to 30 °C, 10% to 75% relative humidity (non-condensing), Installation Category II, Pollution Degree 2
Electrical Specifications 100 Vac to 240 Vac, 50/60 Hz, 300 watts power
Compliance CE Mark; EMC directive (2004/108/EC); Low voltage directive (2006/95/EC); ROHS
Data Output USB and RS-485