POLYVISC®

Sku#

9724-Z22

PolyVISC® High Temperature Dilute Solution Polymer Viscometer


 



In stock

This is non stock item.

PolyVISC is a high temperature, fully automated, benchtop capillary viscometer with an 11 position sample handler for unattended processing and testing of of dilute solution viscosity of polymers in organic solvents. The ultra-stable thermostatic air chamber provides temperature control up to 135 °C. Conforms to ASTM D1601, ASTM D2857, ASTM D4020, ISO 1628-1, ISO 1628-3.

Factory installation of this unit is required. If factory installation is waived, only parts are covered during the warranty period.  Damage occurring from incorrect installation is not covered by warranty.

Designed to meet specific polymer industry needs

  • Meets methods D1601,D2857,D4020,and ISO 1628-1,1628-3
  • On-board software with specialized polymer calculations determines relative,inherent,reduced and intrinsic viscosity
  • Ultra-stable thermostatic air chamber with AIRBATH® technology provides excellent temperature control from 100 °C to 135 °C (± 0.03 °C).
  • Kinematic viscosity range: 0.3 mm2/s (cSt) to 30 mm2/s (cSt). 

Fully automated bench-top testing

  • User-friendly software controls the instrument and facilitates tasks such as sampling,measurement and viscometer tube washing.
  • 11 position sample handler allows for unattended operation.
  • Automated processing reduces operator to operator variability.

Compact,robust design

  • Bench-top design conserves valuable lab space.

Simplified maintenance & test versatility

  • AIRBATH® chamber allows for rapid temperature and viscometer changes.
  • Uses modified Ubbelohde tubes.
  • Volatile sample testing option provides protection from solvent evaporation to enhance test precision and reduce the presence of toxic or corrosive vapors in the AIRBATH® chamber.
Specifications
Method ASTM D1601, ASTM D2857, ASTM D4020, ISO 1628-1, ISO 1628-3
Maximum Throughput 6 tests per hour
Sample Capacity 11
Viscosity Range 0.3 mm²/s (cSt) to 30 mm²/s (cSt)
Timing Resolution 0.01 s
Temperature Range and Accuracy 100 °C to 135 °C ± 0.03 °C
Minimum Sample/Solvent Volume 10 mL sample/20 mL solvent per test
Operating Conditions 15 °C to 30 °C,10% to 75% relative humidity (non-condensing),Installation Category II; Pollution Degree 2
Electrical Specifications 100 Vac,50/60 Hz; 115 Vac,50/60 Hz; 230 Vac,50/60 Hz; 1000 watts power consumption
Compliance CE Mark; EMC directive (2004/108/EC); Low voltage directive (2006/95/EC); HI-POT (1900 Vdc,60 sec.); ROHS
Additional Standard Features Internal heated waste drain lines,dual solvent input
Data Output RS-232 via RS-485