CCS Standards Viscosity Values

CCS Low Temperature Viscosity Table

Low temperature CCS viscosity standards are used for measuring dynamic viscosity (in mPa·s or cP) at temperature from −5 °C to −40 °C. They are used to calibrate the CANNON Cold-Cranking Simulator (CCS) for oil testing per ASTM D5293, Apparent Viscosity of Engine Oils and Base Stocks Between −5 °C and −35 °C Using Cold-Cranking Simulator and SAE specification J300. Lot specific viscosity data in mPa·s (cP) is provided with each standard.

Nominal viscosity values (in mPa·s or cP) of low temperature viscosity CCS standards for ASTM D5293 and SAE specification J300

Catalog Number

Viscosity Standard

−40 °C

−35 °C

−30 °C

−25 °C

−20 °C

−18 °C

−15 °C

−10 °C

−5 °C

9727-N02

CL080

1430

850

9727-N04

CL090

1930

1150

9727-N06

CL100
[CL10]

2450

1450

875

9727-N07

CL110

2880

1680

1030

9727-N08

CL120
[CL12]

3580

2050

1230

775

9727-N09

CL130

4580

2600

1550

975

9727-N10

CL140
[CL14]

6380

3550

2080

1300

775

675

9727-N11

CL150

7830

4310

2500

1530

950

9727-N12

CL160

10300

5580

3200

1900

1200

1000

9727-N13

CL170

12000

6430

3650

2180

1330

850

9727-N14

CL190
[CL19]

15900

8380

4680

2750

1680

1400

1080

9727-N16

CL200

21100

10900

6030

3500

2130

1330

875

9727-N18

CL220
[CL22]

— 

13600

7380

4230

2550

2050

1600

1030

— 

9727-N20

CL240

— 

16900

9100

5180

3050

— 

1900

1230

— 

9727-N22

CL250
[CL25]

— 

20000

10700

6000

3500

2880

2180

1380

— 

9727-N24

CL260

— 

— 

13100

7300

4300

— 

2650

1680

— 

9727-N26

CL280
[CL28]

— 

— 

16500

9080

5280

4300

3200

2030

— 

9727-N28

CL300

— 

— 

20700

11300

6480

— 

3880

2430

— 

9727-N30

CL320
[CL32]

— 

— 

26500

14300

8150

6600

4850

3000

— 

9727-N31

CL340

— 

— 

 —

17000

9580

— 

5650

3480

— 

9727-N32

CL380
[CL38]

— 

— 

 —

20800

11600

9330

6800

4180

2650

9727-N33

CL420

— 

— 

 —

— 

14000

— 

8180

4950

— 

9727-N34

CL480
[CL48]

— 

— 

 —

— 

17400

13700

10100

6000

3780

9727-N35

CL530

— 

— 

 —

— 

22000

— 

12500

7500

— 

9727-N36

CL600
[CL60]

— 

— 

 —

— 

— 

20300

15700

9300

5700

9727-N37

CL680

— 

— 

 —

— 

— 

— 

19400

11300

— 

9727-N38

CL740
[CL74]

— 

— 

 —

— 

— 

— 

— 

12600

7650

9727-N43

Set of 18

Set of CCS-2050/2100 low temp standards (includes CL080, CL090, CL100, CL110, CL120, CL130, CL140, CL150, CL160, CL190, CL220, CL250(2), CL280, CL320, CL380, CL480 and CL600)

This table presents nominal data. Actual values, provided with each standard, may vary slightly from lot to lot and formulations are subject to change. If your application requires close conformity with published nominal values, please contact technical service to obtain the actual viscosity values associated with the current formulation.