High Viscosity Table

High Viscosity Table

Nominal viscosity values of high viscosity standards

This table presents nominal data. Actual values, provided with each standard, may vary slightly from lot to lot and formulations are subject to change. If your application requires close conformity with published nominal values, please contact technical service to obtain the actual viscosity values associated with the current formulation

Catalog Number

 Viscosity Standard

Dynamic viscosity in mPa·s (cP)

Kinematic viscosity mm2/s (cSt)

20 °C
68 °F

25 °C
77 °F

40 °C
104 °F

60 °C
140 °F

135 °C
275 °F

60 °C
140 °F

100 °C
212 °F

135 °C
275 °F

9727-E10

N600*

—

1100

—

160

15

200

—

22

9727-C52

N1000†*

2900

2000

800

280

—

350

80

—

9727-E15

N2000*

—

4900

—

380

21

440

—

26

9727-C57

N4000†*

17000

11000

2900

730

—

850

120

—

9727-E20

N8000*

—

20000

—

1400

—

1600

—

—

9727-C62

N15000†*

68000

41000

11000

2600

—

3000

360

—

9727-E25

N30000*

120000

70000

—

4700

—

5400

—

—

9727-E27

N62000

—

200000

—

13000

—

—

1600

—

9727-E29

N150000

—

420000

—

24000

—

—

3000

—

9727-E30

N190000

840000

520000

140000

33000

—

—

4000

—

9727-E35

N450000

—

1600000

—

100000

2350

—

—

—

9727-E40

N2700000

—

5300000

—

340000

—

—

—

—

9727-E42

N2700000SP

 Dynamic shear rheometer (DSR) standard at 52 °C, 58 °C, 64 °C, 70 °C and 76 °C.

† Viscosity data for additional temperatures provided with these standards are listed in General Viscosity tables.

* Density values provided at all test temperatures for indicated standard