ERAFLASH is preconfigured for the automatic measurement in full compliance with the latest and safest Continuously Closed Cup Flash Point (CCCFP) standards ASTM D6450 and ASTM D7094 and in excellent correlation with ASTM D93, ASTM D56, ASTM D3828 ISO 2719, ISO 13736, EN ISO 3679 and 3680.

The innovative CCCFP methods use only 1 mL to 2 mL of sample that is sealed in a closed chamber during the whole measurement process. An electrical arc inside the chamber ignites the sample and the lack of oxygen extinguishes the flame automatically. The device measures the corresponding pressure increase inside the chamber.

Specifications
Unit Weight 9.7 kg (21.4 lb)
Method ASTM D396, ASTM D975, ASTM D2880, ASTM D3699, ASTM D6450, ASTM D7094, ASTM D7467
Dimensions (W x D x H) 29 cm × 35 cm × 34 cm (11.4 in × 13.8 in × 13.4 in)
Maximum Throughput Up to 12 samples per hour
Temperature Range and Accuracy

0 °C to 200 °C (32 °F to 390 °F) stand alone. No external chiller required.
Down to -25 °C (-13 °F) with water cooling/chiller
Up to 420 °C (788 °F) with optional High Temperature Module 
-40 °C to 120 °C (-40 °F to 248 °F) EREFLASH LT Model with external cooling

Temperature stability: 0.1 °C (0.2 °F)

Minimum Sample/Solvent Volume 1 mL (ASTM D6450) or 2 mL (ASTM D7094)
Electrical Specifications

Auto-switching 85 - 264 V AC,47 - 63 Hz max. 150 W (multi-voltage power supply)
Field application: 12 V DC (vehicle battery) adaptor available

Spectral Libraries Over 100,000 detailed test reports stored in internal memory