TE-DPV® Asphalt

Sku#

9725-F72

TE-DPV® Asphalt Thermoelectric Rotational Paddle Viscometer

In stock

TE-DPV Asphalt is an automated, thermoelectrically cooled rotational digital dual-paddle viscometer for measuring the dynamic viscosity of emulsified asphalts at 25 °C, 40 °C, 50 °C, 60 °C, and 70 °C. The TE-DPV Asphalt provides a convenient, automated alternative to older, labor-intensive, and less precise methods.

Conforms to test methods ASTM D7226, ASTM D2397, ASTM D977, and AASHTO T 382.

Common applications are emulsifited asphalts, maine fuels, suspensions, slurries, and residual oils.

Designed specifically for viscosity testing of emulsified asphalt

  • Meets ASTM D7226, D2397, D977, and AASHTO TP 382 for asphalt emulsions
  • Viscosity range: 30 mPa·s (cP) to 3000 mPa·s (cP); ±5% accuracy
  • Integrated heating and cooling enables dynamic viscosity measurement at 25 °C,40 °C,50 °C,60 °C and 70 °C (± 0.1 °C); CANNON calibrated at 50 °C
  • Digital display presents viscosity values in mPa·s (cP) as well as mm²/s (cSt) for samples with known density values

Improves data quality

  • Automated sample processing and testing reduces the impact of operator on data quality
  • Dual paddles continuously mix the sample for greater homogeneity and more representative data
  • Direct temperature measurement of sample, rather than the bath, improves measurement accuracy and results
  • Viscosity data is automatically calculated and may be transferred to an optional dot matrix printer

Eliminates laborious and tedious methodologies

  • Simply load the sample and push a button to initiate testing
  • Removable sample cup makes cleaning between samples quicker and easier
  • Unattended sample processing saves time and labor
  • Completes most tests in 10 to 30 minutes

Reduces safety risks

  • No hot dripping sample; no glass components or accessories
  • Wipe-clean paddles and sample cup reduce the need to use hazardous cleaning solvents
  • Heater cup thermostat senses over-temperature fault conditions and removes power until temperature reaches a safe level
  • Motor shaft and paddle immobilization automatically cut off motor power

Low cost, easily transportable design

  • Self-contained, compact instrument with no ancillary equipment
  • At 1/7 the benchtop footprint and 1/3 the weight of traditional Saybolt viscometer baths, the TE-DPV Asphalt conserves valuable lab space and transports easily for field applications
Specifications
Unit Weight 10.0 kg (22.0 lb)
Method ASTM D977, ASTM D2397, ASTM D7226, AASHTO T 382
Dimensions (W x D x H) 23.5 cm × 27.3 cm × 45.1 cm (9.3 in × 10.8 in × 17.8 in); 15 cm (6 in) rear clearance required
Shipping Dimensions (W x D x H) 38.1 cm × 40.6 cm × 53.3 cm (15.0 in × 16.0 in × 21.0 in)
Shipping Weight (With All Items) 16.3 kg (36.0 lb)
Maximum Throughput 2 to 4 samples per hour
Sample Capacity 1
Viscosity Range 30 to 3000 mPa·s (CP); ±5%
Temperature Range and Accuracy 25 °C, 40 °C, 50 °C, 60 °C, 70 °C; ±0.1 °C; factory calibrated at 50 °C
Minimum Sample/Solvent Volume 135 mL per test
Operating Conditions 15 °C to 30 °C, 10% to 75% relative humidity (non-condensing), Installation Category II, Pollution Degree 2
Electrical Specifications 100 Vac to 240 Vac, 47 to 63 Hz, 120 watts power consumption
Compliance CE Mark; EMC directive (2004/108/EC); Low voltage directive (2006/95/EC); HI-POT (1900 Vdc,60 sec.); ROHS
Data Output RS-232C